Televac® MC300 Vacuum Controller

Discontinued: The direct replacement is the MX200 Vacuum Controller.

Part Numbers: 2-4502-301, 2-4502-302, 2-4502-303, 2-4502-304, 2-4502-305, 2-4502-306, 2-4503-401, 2-4503-402


Compatible With: 1E Piezo Diaphragm, 1F Piezo Diaphragm, 2A Thermocouple, 4A Convection, 7B Cold Cathode, 7E Cold Cathode, 7F Cold Cathode, 7FC Cold Cathode, 7FCS Cold Cathode, 3F Hot Ion


Caratteristiche e vantaggi:

  • A prezzi competitivi
  • Autodiagnosi integrata
  • Facile da leggere, grande display a LED
  • Quattro relè di processo con set point assegnabili individualmente
  • Uscite analogiche per ogni sensore
  • Eccellente assistenza clienti

The Televac® MC300 vacuum controller controls up to two Televac® thermocouple gauges or convection gauges and an optional cold cathode gauge or mini BA hot ion high vacuum gauge. It features a 0 to 10 V DC analog output for each vacuum gauge, 4 relay set points for process control, and large, bright LED display, making it easy to read from a distance.

Note: sensors and cables are not included and must be ordered separately.

Campo di funzionamento 1*10-11 Torr to 1000 Torr
Comunicazioni Nessuno
Uscita analogica 0 ... 10 V DC (1 per sensore)
Risoluzione di uscita analogica 12 bits
Display Output Resolution 3 significant digits
Set point programmabili 4
Tipo di set point Relè
Tensione di alimentazione 115/230 V AC (0.3/0.15 A), 50-60 Hz
Potenza massima 35 W
Mezzo di calibrazione Aria secca o azoto
Sovrappressione Dipendente dal sensore
Temperatura di esercizio 0 °C a 50 °C
Temperatura di stoccaggio da -40 °C a 85 °C
Distanza leggibile del display 10 m (35 ft)
Sensori di massimo controllo Up to 3

The MC300 has four relay set points. Individual relay status LEDs are displayed on the front panel. The heavy duty relays have a SPDT (1 Form C) contact and are rated for 24 V DC. They have a maximum current rating of 8 A at 250 V AC or 5 A at 30 V DC. The data sheet can be found at http://www.omron.com/ecb/products/pdf/en-g6rn.pdf.

The MC300 is sold in configurations containing high vacuum sensors (cold cathode or hot ion). The high vacuum sensor is controlled automatically by the low vacuum sensor connected to station 1. It is switched on when station 1 reads below 10 microns (10 mTorr or 0.010 Torr). When operating high vacuum sensors, it is recommended that the channel 3 high vacuum sensor and channel 1 low vacuum sensor be located near one another on the same manifold.

Alloggiamento Cold rolled steel
Collegamenti elettrici Power cord (IEC 60320 C-13)
Weight (Base Unit) 2.2 kg (5 lbs)
Mounting Panel mount or half rack adaptable

Dimensional Drawing of the Televac® MC300 Vacuum Controller

Certified to UL 61010-1
CE certified to EN 61326-1:2006
Certified to CAN/CSA C22.2 No. 61010-1-12
ISO 17025 accredited calibration optional
NIST-traceable calibration optional
Controllore di vuoto MC300 Numero di parte
Dual 2A with 7B (10-7 to 20 Torr) 2-4502-301
Dual 4A with 7B (10-7 to 1000 Torr 2-4502-302
Dual 2A with 7E/7F/7FC (10-11 to 20 Torr) 2-4502-303
Dual 4A with 7E/7F/7FC (10-11 to 1000 Torr) 2-4502-304
Dual 2A with Mini BA (10-10 to 20 Torr) 2-4502-305
Dual 4A with Mini BA (10-10 to 1000 Torr) 2-4502-306
Dual 2A (10-3 to 20 Torr) 2-4503-401
Dual 4A (10-3 to 1000 Torr) 2-4503-402